Cardiac Science · Model Quinton ST55
Features digitally controlled AC-drive system that delivers greater performance, improved accuracy, and increased reliability. Speed start at.8 mph to 9.6 mph. 400 lb user weight capacity. Length: 80.2” (204cm). 115V/20A power required. Dedicated circuit required.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Belt will not run, no display on the ECG monitor, no response to start.
Root cause: Building or treadmill circuit breaker tripped, blown treadmill fuse, a cut power cord, or a faulty control cable between the variable-speed drive (VSD) and the ECG monitor, or a VSD failure.
Symptoms: Belt speed or incline does not match the setting, error code on the ECG monitor, grade drifts.
Root cause: Grade potentiometer needs calibration, a grade-motor / limit-switch fault, or a VSD speed-feedback fault.
Symptoms: Belt slips, drifts off-centre, rubs the frame, rough or noisy running.
Root cause: Worn or stretched walking belt, worn deck, or drive-belt tension loss.
Symptoms: Rumble, squeal, or vibration under load.
Root cause: Worn rollers/bearings, drive belt, or motor.
Symptoms: Loose handrail, emergency stop not stopping the belt (a recall flagged inadequate electrical-safety testing on some 2006-2009 units).
Root cause: Loosened fasteners or a failed e-stop switch/wiring.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Walking belt | Periodic replacement | $$ | Moderate |
| Deck | Periodic replacement | $$ | Moderate |
| Drive belt | Periodic replacement | $ | Moderate |
| Grade motor / potentiometer | Major service part | $$ | Moderate |
| Roller / bearing set | Major service part | $$ | Moderate |
| Treadmill fuse | Major service part | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
28 stress and rehab treadmills manufactured between September 2006 and January 2009 may not have received adequate electrical safety testing prior to being shipped from the factory. If these components break down in the field, a patient or operator could be exposed to excess touch currents on the surface of the product. Because safety testing may not have allowed for complete evaluation of the el
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio