With the PROCARE* Monitor B20 from GE Healthcare, there are no compromises. Youll find what youre looking for to monitor your patients. Versatile clinical capabilities to monitor a wide range of patients. Ease of operation for all professional levels. Excellent reliability driven by high quality GE design and engineering. And importantly, the affordability that enables you to provide patients with world-class monitoring technology while staying within your budget The B20 monitor is designed with advanced clinical measurement technologies for accurate and reliable patient monitoring under challenging conditions. Key capabilities include. GE EK-Pro* arrhythmia program with ECG multi-leads simultaneous analysis to optimise the detection of arrhythmias and other cardiac events. GE DINAMAP* SuperSTAT* NIBP for fast, comfortable non-invasive blood pressure readings with a high degree of accuracy. GE SpO2 for reliable, cost-effective oxygen saturation monitoring (Nellcor* OxiMax* SpO2 and Masimo* SET* SpO2 optional). GE Datex-Ohmeda* compact CO2 sidestream respiratory monitoring with a new generation of infrared technology1.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: ECG, IBP, SpO2, temperature, CO2 or CO intermittently drops or is implausible.
Root cause: Worn patient cable / sensor, or a parameter-module / input-board fault.
Symptoms: NIBP times out, cuff will not inflate / hold, readings implausible.
Root cause: Internal NIBP pump or valve, a cuff / hose leak, or a pressure-transducer fault.
Symptoms: The host does not display a plugged-in module or a built-in parameter.
Root cause: Docking / connector fault or an internal board fault.
Symptoms: Blank, dim or distorted display; unresponsive touchscreen.
Root cause: Display, backlight, or controller fault.
Symptoms: Short battery run time, will not hold charge, will not power on.
Root cause: Aged battery pack or a power-supply fault.
Symptoms: Monitor not shown at the central station, admit fails, intermittent drop.
Root cause: Wired / Wi-Fi network configuration or a communication board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable set (ECG / IBP / SpO2 / temp) | Frequent wear item | $$ | Easy |
| NIBP cuff and hose | Frequent wear item | $ | Easy |
| Internal NIBP pump / valve | Major service part | $$$ | Difficult |
| Parameter module / input board | Major service part | $$$ | Difficult |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Display / touchscreen assembly | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.