GE HealthCare · Model Marquette Case 16
12 SL Stress Test with Treadmill Stroke-written display exposes waveform variations User-defined measurements assures accuracy.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Baseline wander, muscle / motion noise, "lead fail", ST measurements unreliable.
Root cause: Poor skin prep, dried or lifting electrodes, a worn acquisition (CAM / patient) cable, or EMI from the treadmill.
Symptoms: CASE cannot control the treadmill or bike, protocol will not advance, speed / grade error.
Root cause: Treadmill control cable / interface, or a treadmill drive / grade fault.
Symptoms: Channels drop, the CASE application hangs, test will not start.
Root cause: Acquisition (CAM) module or the workstation PC.
Symptoms: Exercise BP times out or is implausible.
Root cause: NIBP pump / valve or a cuff / hose leak.
Symptoms: No printed report, or reports will not send to MUSE / the EMR.
Root cause: Printer, or the network configuration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Acquisition (CAM) module / patient cable | Periodic replacement | $$$ | Fairly easy |
| Treadmill control cable / interface | Periodic replacement | $$ | Fairly easy |
| Workstation PC | Major service part | $$$ | Moderate |
| NIBP cuff / hose (if fitted) | Frequent wear item | $ | Easy |
| Report printer | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio