GE HealthCare · Model Dash 5000
Design for optimized monitoring in all care applications. Advanced algorithms for all monitored parameters. Fully flexible alarm settings allow personalized monitoring. ECG acquisition through cables or wireless. Bright alarm light at the bedside helps to immediately locate the patient in critical situations. Multiple mounting solutions and a wide range of accessories ensure easy and flexible workplace configuration. 3 waveform screen view to maximize the size of parameter values and waveforms. Convenient care area specific Remote iTouch Pad enables operation on both sides of the patient’s bed and direct access to frequently used functions. Remote Screen through DashPort2 for optimal central overview. CO2 module and extention Rac for additional parameters. Wide array of interfacing options with iConnect solutions. Both mainstream and low-flow sidestream CO2 measurement options for applications from neonatal to adult patients Optimized for demanding transportation. Light weight, durable, and ergonomic design. Up to 2 Hot Swap batteries for unlimited battery powered monitoring. DashPort2 Docking Station for easy switching between bedside and transport monitoring. Continuous central monitoring during transport through a Wireless Network. Compatible with bed mount for in-hospital transport.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: ECG, IBP, SpO2, temperature, CO2 or CO intermittently drops or is implausible.
Root cause: Worn patient cable / sensor, or a parameter-module / input-board fault.
Symptoms: NIBP times out, cuff will not inflate / hold, readings implausible.
Root cause: Internal NIBP pump or valve, a cuff / hose leak, or a pressure-transducer fault.
Symptoms: The host does not display a plugged-in module or a built-in parameter.
Root cause: Docking / connector fault or an internal board fault.
Symptoms: Blank, dim or distorted display; unresponsive touchscreen.
Root cause: Display, backlight, or controller fault.
Symptoms: Short battery run time, will not hold charge, will not power on.
Root cause: Aged battery pack or a power-supply fault.
Symptoms: Monitor not shown at the central station, admit fails, intermittent drop.
Root cause: Wired / Wi-Fi network configuration or a communication board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable set (ECG / IBP / SpO2 / temp) | Frequent wear item | $$ | Easy |
| NIBP cuff and hose | Frequent wear item | $ | Easy |
| Internal NIBP pump / valve | Major service part | $$$ | Difficult |
| Parameter module / input board | Major service part | $$$ | Difficult |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Display / touchscreen assembly | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.