GE HealthCare · Model ApexPro CH Telemetry System
ApexPro CH provides exceptional detection and analysis of patient data to help you fully and accurately monitor your patients’ status. ApexPro CH uses the GE EK-Pro clinical algorithm, which processes and analyzes four independent, simultaneous ECG leads. Plus, an innovative Smart Leads Fail feature helps to reduce false ECG alarms. With GE EK-Pro, ApexPro CH supports advanced atrial fibrillation detection and alarming. ApexPro CH has the capability to monitor two V-leads on each patient. Pace detection across two vectors improves the system’s ability to see and recognize a wide variety of pacemakers. The flexible design of ApexPro CH allows each transmitter to monitor only the parameters each patient requires. Continuous monitoring of ECG and SpO2, and episodic monitoring of NIBP and temperature allow caregivers to tailor monitoring according to the patient’s acuity.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: ECG, IBP, SpO2, temperature, CO2 or CO intermittently drops or is implausible.
Root cause: Worn patient cable / sensor, or a parameter-module / input-board fault.
Symptoms: NIBP times out, cuff will not inflate / hold, readings implausible.
Root cause: Internal NIBP pump or valve, a cuff / hose leak, or a pressure-transducer fault.
Symptoms: The host does not display a plugged-in module or a built-in parameter.
Root cause: Docking / connector fault or an internal board fault.
Symptoms: Blank, dim or distorted display; unresponsive touchscreen.
Root cause: Display, backlight, or controller fault.
Symptoms: Short battery run time, will not hold charge, will not power on.
Root cause: Aged battery pack or a power-supply fault.
Symptoms: Monitor not shown at the central station, admit fails, intermittent drop.
Root cause: Wired / Wi-Fi network configuration or a communication board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable set (ECG / IBP / SpO2 / temp) | Frequent wear item | $$ | Easy |
| NIBP cuff and hose | Frequent wear item | $ | Easy |
| Internal NIBP pump / valve | Major service part | $$$ | Difficult |
| Parameter module / input board | Major service part | $$$ | Difficult |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Display / touchscreen assembly | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Intermittent or continuous loss of patient monitoring.
When a patient being monitored with ApexPro/Apex Pro CH Telemetry with ST monitoring disabled is in a pre-existing condition of continuous MESSAGE or ADVISORY level alarm preceding a SYSTEM WARNING level alarm, the SYSTEM WARNING audible alarm and flashing yellow border around the patient panel at the CIC (Clinical Information Center) does not occur.
System Warning Alarm failure: When a patient being monitored is in a pre-existing condition of continuous MESSAGE or ADVISORY level alarm preceding a SYSTEM WARNING level alarm, the SYSTEM WARNING audible alarm and flashing yellow border around the patient panel at the CIC does not occur.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio