Fanem · Model Heater Box 1502
Digital control panel with temperate settings from +5°C of the environment to 60°C, with 0.1°C resolution Thermostatic accuracy from +/- 0.7°C at 37°C Temperature preset at 37°C, controlled by a digital sensor, with internal circulation of heated air generated by a tangential fan with a capacity of 150 m3/h in order to ensure better uniformity Uniformity of +/- 1.5°C, at a temperature of 37°C Easy-to-replace, modular heating element Electronic safety thermostat with a chip that is separate from the control system; automatically shuts down the heating system and issues a visual warning on the panel Control panel with LEDs which visually indicate the current temperature and heating, with increase and decrease keys and a key for setting parameters High temperature digital alarm that is totally separate from the temperature control, with audible and visual indication Safety switch which stops air circulation and heating when the lid is open and re-establishes normal operation when the lid is closed.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Output below target, low-temperature alarm, slow warm-up.
Root cause: Heater plate(s) or the heater-control circuit, or a temperature-sensor fault.
Symptoms: The unit disables heating with an over-temperature condition (protects against haemolysis).
Root cause: Sensor or control fault driving the heater high, or an extreme flow-rate change.
Symptoms: Air-in-line alarm, flow stops.
Root cause: Air in the set, an improperly primed set, or an air-detector fault.
Symptoms: Set not detected, will not clamp / seat, leaks at the warming chamber.
Root cause: Worn set-detection switch or chamber fitting.
Symptoms: Cannot reach set flow, pressure alarm.
Root cause: Pump / roller fault or an occlusion.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Heater plate(s) / heater-control assembly | Periodic replacement | $$$ | Moderate |
| Temperature sensor | Major service part | $$ | Moderate |
| Air-detector / bubble sensor | Major service part | $$ | Moderate |
| Set-detection switch / chamber fitting | Periodic replacement | $$ | Fairly easy |
| Disposable warming set | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio