Fanem · Model Heamto-Immuno EBS 3347
Automatic defrosting system, with evaporation of condensation Condensation tank is removable to enable sterilization in accordance with quality and safety standards Forced air circulation results in perfect homogenization within the chamber Fan operates even when batteries are used, ensuring uniformity in any system Easy-to-view LCD display and real-time LED indicators of the status of the auxiliary battery Digital thermometer with 0.1oC subdivision of readings, and maximum and minimum temperatures read directly and simultaneously on the same display Menu for multiple sensors, which enables simultaneous viewing of the temperature in all installed sensors Audible and visual alarms for open door, pre-programmed low and high temperatures and EBS status at 2oC and 8oC, respectively; can be adjusted manually at other temperatures Alarm testing system for simulation of low and high temperature alarms Operating temperature: 2oC to 8oC, memorized at 4oC Internal capacity: 120 liters/9,100 5-ml doses DAQ (Data Acquisition) - registers data in memory for up to 10 days, allowing for subsequent consultation of temperature variations.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: No heat, overheat alarm.
Root cause: Element or thermal fuse failure.
Symptoms: Setpoint error, alarm.
Root cause: Sensor aging.
Symptoms: Uneven warming, hotspot.
Root cause: Blower wear.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable per patient.
Common R&R.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Temperature probe Consumable per patient. | Periodic replacement | $ | Easy |
| Heating element / thermal fuse Common R&R. | Periodic replacement | $$$ | Moderate |
| Blower fan | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →