Estill Medical Technologies · Model Thermal Angel TA-200
• Lightweight (9 oz), in-line, disposable, portable battery-powered blood and IV fluid infusion warmer • No cartridges or cleaning • Inserted below standard IV tubing, coming from the bag and before the extension set leading to the infusion site • Quick setup within 30 seconds • Warms in approximately 45 seconds • Placed at or near the infusion site for minimal heat loss through tubing • Fail-safe to prevent overheating • Ability to remain with patient from first contact through hospital stay • Can be used in the field, ambulances, helicopters, ER, OR, military and other care areas • The unit uses standard luer fittings and will accept all standard IV line sets • Single patient use • Heaters adjust automatically in response to changes in flow rate • Innovative Technology
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: No heat, or overheat alarm.
Root cause: Heater element or thermal cutoff failure.
Symptoms: Delivered fluid temperature out of set range.
Root cause: Inline/probe sensor aging or miscalibration.
Symptoms: False or missed air-in-line alerts.
Root cause: Air detector sensor drift or dirty optics.
Symptoms: Fluid leak, or won't prime.
Root cause: Worn cassette/tubing seat or cracked fitting.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable in some designs.
Safety-critical; verify at PM.
Single-use consumable.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Heating element / plate | Periodic replacement | $$$ | Moderate |
| Temperature sensor / probe Consumable in some designs. | Periodic replacement | $$ | Fairly easy |
| Air-in-line sensor Safety-critical; verify at PM. | Major service part | $$$ | Moderate |
| Disposable tubing / cassette set Single-use consumable. | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Estill has discovered that a fully discharged Ultra Battery 1 (TA-UB1) will reach 96.9% of its capacity during a 60 minute charging period. If you have a fully discharged Ultra Battery 1 (TA-UB1), in order to achieve a 100% of charge capacity you should charge it for a period of 90 minutes. In addition, the charging voltage range of 10-36V specified on the Ultra Battery 1 (TA-UB1) label and in t
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.