Ergoline · Model Ergoselect 400
Arm ergometer ergoselect 400 The perfect choice for the performance of exercise tests with disabled patients or wheelchair users. The special chair, which is permanently attached to the ergometer, and the anti-slip footrests offer a high level of patient comfort and safety.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Arm-crank resistance wrong or absent, load error during testing.
Root cause: Eddy-current brake or control board calibration, or speed-feedback fault.
Symptoms: Rumble or uneven resistance at the hand crank.
Root cause: Worn drive belt or crank bearing.
Symptoms: Exercise system does not see the ergometer.
Root cause: Interface cable or control board.
Symptoms: Fixed patient chair or crank-height adjustment will not lock.
Root cause: Worn clamp or adjustment hardware on the wheelchair-accessible frame.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Eddy-current brake unit | Major service part | $$$ | Difficult |
| Drive belt | Periodic replacement | $ | Moderate |
| Crank bearing set | Major service part | $$ | Difficult |
| Control PCB | Major service part | $$$ | Moderate |
| Chair / height clamp kit | Major service part | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.