EllTec · Model Animec AM-4
Automatic Temperature Control - Heating is controlled by one effluent temperature sensor and two heating plate temperature sensors. Temperature Display - Easy to read by LED display shows the temperature of the fluid. Accurate fluid temperature increments of 0.1 ºC degrees. Alarm - An audible buzzer and visual alarm are activated at a fluid temperature of 41.5 ºC (AM-412), 39 ºC (AM-413). The alarm buzzer volume is adjustable. Quick set-up - The L-fusion set is placed directly between the heating plates. Rapid warm-up - Ready to use in 2~3 minutes. Safe performance - Protection from over-heating, electric shock and contamination Protection from air bubbles is provided via the air chamber in the L-fusion disposable set. High durability - Simple construction and shock resistant.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Output below target, low-temperature alarm, slow warm-up.
Root cause: Heater plate(s) or the heater-control circuit, or a temperature-sensor fault.
Symptoms: The unit disables heating with an over-temperature condition (protects against haemolysis).
Root cause: Sensor or control fault driving the heater high, or an extreme flow-rate change.
Symptoms: Air-in-line alarm, flow stops.
Root cause: Air in the set, an improperly primed set, or an air-detector fault.
Symptoms: Set not detected, will not clamp / seat, leaks at the warming chamber.
Root cause: Worn set-detection switch or chamber fitting.
Symptoms: Cannot reach set flow, pressure alarm.
Root cause: Pump / roller fault or an occlusion.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Heater plate(s) / heater-control assembly | Periodic replacement | $$$ | Moderate |
| Temperature sensor | Major service part | $$ | Moderate |
| Air-detector / bubble sensor | Major service part | $$ | Moderate |
| Set-detection switch / chamber fitting | Periodic replacement | $$ | Fairly easy |
| Disposable warming set | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.