EllTec · Model Animec AM-2S
• Automatic regulation - The temperature of the warming plate adjusts according to the desired temperature of the effluent. • No disposables - Special warming bags or coils are not required. • Less heat loss after warming - Small size neccessitates minimal tubing between the patient and the unit preventing cooling of the infused fluid. • Quick set-up - The tube is inserted directly into the warmer; no priming or disposable preparation is required. • Rapid warm-up - Ready to use in 2 ~ 3 minutes • Safe performance - Protection from overheating, electric shock and contamination. • High durability - Simple construction and shock resistant casing.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: No heat, or overheat alarm.
Root cause: Heater element or thermal cutoff failure.
Symptoms: Delivered fluid temperature out of set range.
Root cause: Inline/probe sensor aging or miscalibration.
Symptoms: False or missed air-in-line alerts.
Root cause: Air detector sensor drift or dirty optics.
Symptoms: Fluid leak, or won't prime.
Root cause: Worn cassette/tubing seat or cracked fitting.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable in some designs.
Safety-critical; verify at PM.
Single-use consumable.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Heating element / plate | Periodic replacement | $$$ | Moderate |
| Temperature sensor / probe Consumable in some designs. | Periodic replacement | $$ | Fairly easy |
| Air-in-line sensor Safety-critical; verify at PM. | Major service part | $$$ | Moderate |
| Disposable tubing / cassette set Single-use consumable. | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →