Resource management benefits Caseload versatility Workflow efficiency Patient throughput Treatment quality benefits Patient safety Targeting accuracy Dose escalation Hypofractionation A wide range of stereotactic and intensity modulated treatments Lung Liver Prostate Head-and-neck Spinal metastases Large brain metastases.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Beam-on faults, dose-rate low or unstable, RF/HV or arc interlocks.
Root cause: Magnetron or thyratron at end of life, HV supply, or waveguide arcing / SF6 pressure loss.
Symptoms: Leaf stuck or out of position, delivery interlock, QA fluence errors.
Root cause: MLCi / Agility leaf motor, leaf-guide wear, or a leaf position-readout fault.
Symptoms: Rotation or field-size error, position interlock.
Root cause: Drive motor, encoder, or brake on the gantry or collimator.
Symptoms: kV source or flat-panel imaging degraded, poor CBCT for setup, panel/arm deploy errors.
Root cause: kV X-ray tube, amorphous-silicon panel, or the imaging-arm drive.
Symptoms: Output constancy checks trend out, symmetry/flatness off, chamber interlock.
Root cause: Monitor ion-chamber aging or a dose-servo fault.
Symptoms: Over-temperature interlocks, dose-rate instability, coolant-flow alarm.
Root cause: Chiller pump, heat-exchanger fouling, or low coolant.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
OEM service; dosimetry re-commissioning required.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Magnetron OEM service; dosimetry re-commissioning required. | Periodic replacement | $$$$ | Expert-level |
| Thyratron | Periodic replacement | $$$$ | Expert-level |
| MLC leaf motor / drive assembly | Major service part | $$$$ | Expert-level |
| kV X-ray tube / a-Si imaging panel | Major service part | $$$$ | Expert-level |
| Monitor ion chamber | Major service part | $$$$ | Expert-level |
| Chiller pump / heat exchanger | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.