Siemens Healthineers · Model ONCOR Avant-Garde
Gated Therapy allows for a seamless delivery process OPTIFOCUS Multileaf Collimator for accurate and flexible beam shaping Optimized treatment for better outcomes Advanced user interface for intuitive setup and delivery control Insulated structure reduces noise for patients’ comfort Compact design makes more room in the treatment room A unique portal imaging application for precise and efficient patient positioning All digital platform offers convenience and flexibility A leading-edge treatment table that will hold even the heaviest patients – up to 550 pounds/250 kg.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Beam-on faults, dose-rate low or unstable, RF/HV or arc interlocks.
Root cause: Magnetron/klystron or thyratron at end of life, HV supply, or waveguide arcing / SF6 pressure loss.
Symptoms: Leaf stuck or out of position, delivery interlock, QA fluence errors.
Root cause: MLC leaf motor, leaf-guide wear, or a leaf position-readout fault.
Symptoms: Rotation, field-size or couch position error, position interlock.
Root cause: Drive motor, encoder, or brake.
Symptoms: kV source or flat-panel imaging degraded, poor image for setup, imaging-arm deploy errors.
Root cause: kV X-ray tube, amorphous-silicon panel, or the imaging-arm drive.
Symptoms: Output constancy checks trend out, symmetry/flatness off, chamber interlock.
Root cause: Monitor ion-chamber aging or a dose-servo fault.
Symptoms: Over-temperature interlocks, dose-rate instability, coolant-flow alarm.
Root cause: Chiller pump, heat-exchanger fouling, or low coolant.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
OEM service; dosimetry re-commissioning required.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Magnetron / klystron OEM service; dosimetry re-commissioning required. | Periodic replacement | $$$$ | Expert-level |
| Thyratron | Periodic replacement | $$$$ | Expert-level |
| MLC leaf motor / drive assembly | Major service part | $$$$ | Expert-level |
| kV X-ray tube / a-Si imaging panel | Major service part | $$$$ | Expert-level |
| Monitor ion chamber | Major service part | $$$$ | Expert-level |
| Chiller pump / heat exchanger | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.