Dräger · Model Vista 120
Essential monitoring capabilities. Available in three models, the Vista 120 displays 8 waveforms in an easy-to-configure layout. A core set of essential parameters comes standard on all models: 3/5 lead ECG, SpO2, non-invasive blood pressure, respiration and dual temperature. In addition, advanced parameters are available, including two invasive blood pressures and etCO2, depending on the model. Supports workflow efficiently and cost-effectively. The Vista 120 is easy to learn and easy to use. The bright 380 mm (15") screen provides a clear view of patient data, even from a distance. You can configure the display to see the information you want to see, the way you want to see it. Fast access keys and simplified menus put the data you need right at your fingertips. A nurse call interface allows connection to most hospital call systems. For workflow flexibility, the Vista 120 can be mounted on a ventilator, to an anesthesia machine, or on the wall. Comprehensive set of accessories. Each Vista 120 monitor includes a starter set of accessories for adult patients. A wide range of accessories, including those that support pediatric and neonatal patients, are available separately as options. Part of a complete departmental solution. The Vista 120 supports adult, pediatric and neonatal patients in a variety of care environments – including Intensive Care Units, Operating Rooms, Emergency Departments and Neonatal Intensive Care Units. Use the Vista 120 on its own or with a complementary Dräger device, such as a ventilator, anesthesia machine or baby warmer. The Vista 120 also offers advanced parameters for the ICU, OR, ED and NICU, depending on the model you choose. Built in recorder, power supply and battery. The Vista 120 has an integrated recorder that prints out up to 3 channels of information – saving time by providing documentation when and where you need it. There is a built-in power supply for direct AC connection, as well as an optional 120-minute built-in battery.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: ECG, IBP, SpO2, temperature, CO2 or CO intermittently drops or is implausible.
Root cause: Worn patient cable / sensor, or a parameter-module / input-board fault.
Symptoms: NIBP times out, cuff will not inflate / hold, readings implausible.
Root cause: Internal NIBP pump or valve, a cuff / hose leak, or a pressure-transducer fault.
Symptoms: The host does not display a plugged-in module or a built-in parameter.
Root cause: Docking / connector fault or an internal board fault.
Symptoms: Blank, dim or distorted display; unresponsive touchscreen.
Root cause: Display, backlight, or controller fault.
Symptoms: Short battery run time, will not hold charge, will not power on.
Root cause: Aged battery pack or a power-supply fault.
Symptoms: Monitor not shown at the central station, admit fails, intermittent drop.
Root cause: Wired / Wi-Fi network configuration or a communication board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable set (ECG / IBP / SpO2 / temp) | Frequent wear item | $$ | Easy |
| NIBP cuff and hose | Frequent wear item | $ | Easy |
| Internal NIBP pump / valve | Major service part | $$$ | Difficult |
| Parameter module / input board | Major service part | $$$ | Difficult |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Display / touchscreen assembly | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.