Two-in-One Solution Independent instruments provide flexibility and cost savings to facilities with multiple OR’s, by allowing twin systems to be separated and used as two single units. For IVRA procedures, simply bring unis back together. Intelligent Touch Intuitive user interface simplifies key tourniquet parameters and user training. Personalized Pressure Limits Limits normal maximum pressure to 475 mmHg, and allows selection of extended pressure of up to 600 mmHg if required for individual patients. Advanced Leak Detection During Surgery Monitors the tourniquet cuff, tubing and connectors for air leaks during surgery, and safely alerts staff to hazards. Speed Reduces set-up and inflates a typical 34" thigh cuff in less than 5 seconds. Integrated Tourniquet Cuff Testing Enables cuff testing before or after surgery as recommended in current AORN Standards and Recommended Practices. Advanced Safety Features Intuitive time and pressure safety alarms; unique self-test and calibration test; visible displays of pressure and time. Greatest Battery Life Up to 10 hours of operation with tested cuffs. Connectivity Allows easy integration with OR information networks.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Pressure droops during a case, will not inflate to setpoint, leak alarm.
Root cause: Internal pump / compressor wear, a solenoid-valve fault, or a leak in the tubing / cuff / connectors.
Symptoms: Displayed pressure differs from an independent gauge - patient-safety critical.
Root cause: Pressure-transducer drift or a calibration fault.
Symptoms: One channel will not regulate independently, cross-talk between channels.
Root cause: Channel solenoid / transducer or a manifold fault.
Symptoms: Inflation-time alarm not sounding, timer wrong.
Root cause: Alarm / timer board fault - a safety function.
Symptoms: Leak at the connector, cuff will not hold.
Root cause: Worn tubing, connectors, or a damaged cuff.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Verify against a calibrated gauge after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Internal pump / compressor | Major service part | $$$ | Difficult |
| Solenoid / regulator valve | Major service part | $$ | Moderate |
| Pressure transducer Verify against a calibrated gauge after replacement. | Periodic replacement | $$ | Moderate |
| Connecting tubing / connectors | Frequent wear item | $ | Easy |
| Alarm / control board | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.