Stryker · Model SmartPump
Large LED screen display Minimizes staff training through simple cuff controls Displays and stores reliable, real-time data Includes a 5-hour, internal battery backup Interfaces with printer and external CIS devices.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Pressure droops during a case, will not inflate to setpoint, leak alarm.
Root cause: Internal pump / compressor wear, a solenoid-valve fault, or a leak in the tubing / cuff / connectors.
Symptoms: Displayed pressure differs from an independent gauge - patient-safety critical.
Root cause: Pressure-transducer drift or a calibration fault.
Symptoms: One channel will not regulate independently, cross-talk between channels.
Root cause: Channel solenoid / transducer or a manifold fault.
Symptoms: Inflation-time alarm not sounding, timer wrong.
Root cause: Alarm / timer board fault - a safety function.
Symptoms: Leak at the connector, cuff will not hold.
Root cause: Worn tubing, connectors, or a damaged cuff.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Verify against a calibrated gauge after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Internal pump / compressor | Major service part | $$$ | Difficult |
| Solenoid / regulator valve | Major service part | $$ | Moderate |
| Pressure transducer Verify against a calibrated gauge after replacement. | Periodic replacement | $$ | Moderate |
| Connecting tubing / connectors | Frequent wear item | $ | Easy |
| Alarm / control board | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Synced live from the FDA openFDA API.
Tourniquet cuff flange may become detached from the bladder. If leak occurs during the procedure, there may be risk of hemorrhage, resulting in blood loss greater than expected. Other serious risks include cerebral hypotension and cardiovascular collapse and arrest, risking stroke or death. Specific to IVRA, patient toxicity may also occur due to rapid systemic absorption of anesthetics.
Three complaints have been reported indicating that the product inside the package did not match the label description. The units had quick connectors instead of luer lock connectors as stated in the label description.
This recall is related to the Electi-cord recall. There is a risk when excessive force is applied to the plugs, a fracture can develop on the prongs inside the molded section of the plug. Over time the fracture may lead to arcing inside of the molded sction of the plug and a build up of heat which can cause charring, electric shock or fire. To this date there have been zero complaint for this fa