Symptoms: Pressure droops during a case, will not inflate to setpoint, leak alarm.
Root cause: Internal pump / compressor wear, a solenoid-valve fault, or a leak in the tubing / cuff / connectors.
Symptoms: Displayed pressure differs from an independent gauge - patient-safety critical.
Root cause: Pressure-transducer drift or a calibration fault.
Symptoms: One channel will not regulate independently, cross-talk between channels.
Root cause: Channel solenoid / transducer or a manifold fault.
Symptoms: Inflation-time alarm not sounding, timer wrong.
Root cause: Alarm / timer board fault - a safety function.
Symptoms: Leak at the connector, cuff will not hold.
Root cause: Worn tubing, connectors, or a damaged cuff.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Verify against a calibrated gauge after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Internal pump / compressor | Major service part | $$$ | Difficult |
| Solenoid / regulator valve | Major service part | $$ | Moderate |
| Pressure transducer Verify against a calibrated gauge after replacement. | Periodic replacement | $$ | Moderate |
| Connecting tubing / connectors | Frequent wear item | $ | Easy |
| Alarm / control board | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.