Zimmer · Model ATS 3000
Innovative Limb Occlusion Pressure(LOP) feature. Uses ambient air. Uses electronic and microprocessor regulated pressure. Self-Check Calibration automatically checks the accuracy of the machine calibration each time the unit is powered on. Cuff Alert alerts the user of cuff status when an attempt is made to set the machine on standby. Cuff Lockout feature prompts the user to confirm the deflation command during bilateral or IVRA procedure in order to reduce complications due to sudden cuff deflation of the second cuff.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Main or second cuff does not pressurise on command.
Root cause: Membrane panel not seated, pump not plugged in, a stuck-open deflate valve, or a valve-driver / harness fault.
Symptoms: Cuff stays pressurised when deflation is commanded.
Root cause: Deflate valve stuck shut, or a valve-driver / harness fault.
Symptoms: Displayed pressure disagrees with a reference, self-check on power-up fails.
Root cause: Pressure transducer drift or a control-board calibration fault.
Symptoms: System repeatedly tops up the cuff, leak alarm, set pressure not held.
Root cause: Cuff bladder or hose leak, or a loose positive-locking connector.
Symptoms: Battery indicator lit or flashing during normal use, unit dies on transport.
Root cause: Aged backup battery or a charger fault (battery is for power-loss / transport only).
Symptoms: Unit fails at first use or mid-procedure.
Root cause: Subject of a Zimmer A.T.S. 3000 voluntary removal for out-of-box and intra-operative failures - check the unit's serial against the recall notice.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibrate after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Air pump | Major service part | $$$ | Difficult |
| Inflate / deflate valve assembly | Major service part | $$ | Difficult |
| Pressure transducer Recalibrate after replacement. | Major service part | $$ | Moderate |
| Membrane panel / keypad | Major service part | $$ | Moderate |
| Backup battery | Periodic replacement | $$ | Fairly easy |
| Tourniquet cuff and hose | Periodic replacement | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Synced live from the FDA openFDA API.
Zimmer is conducting a voluntary removal of some Zimmer A.T.S. 3000 tourniquets after receiving complaints of both out-of-box failures and failures during surgical procedures on some of the affected units. During failure the unit will alarm, lock into a non-operational mode, and the display screen will go blank. Based on investigation, this problem appears to be due to a counterfeit chip on the