Delfi · Model PTS
Integrated Tourniquet Cuff Testing-for testing the integrity of the tourniquet cuff with tubing and connectors before or after surgery, as recommended in current AORN standards and recommended practices. Rapid Cuff Inflation & Deflation-Inflates a typical 34″ thigh cuff in less than 5 seconds Extended Pressure Interlock- A safety interlock limits normal maximum pressure to 475mmHg, and allows staff to select an extended pressure of up to 600 mmHg if required for patients with special needs Long Battery Life-Up to 15 hours of operation with tested cuffs Connectivity Allows easy integration with OR information networks Advanced Leak Detection During Surgery-Monitors the tourniquet cuff, tubing and connectors for air leaks during surgery, and safely alerts staff to hazards or follow-up actions required Color Display & Intuitive User Interface-For clearly visible key tourniquet parameters and simplified operator training Printer{Optional} Provides a record of events including tourniquet time and pressure levels, for easy inclusion in patient records.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Pressure droops during a case, will not inflate to setpoint, leak alarm.
Root cause: Internal pump / compressor wear, a solenoid-valve fault, or a leak in the tubing / cuff / connectors.
Symptoms: Displayed pressure differs from an independent gauge - patient-safety critical.
Root cause: Pressure-transducer drift or a calibration fault.
Symptoms: One channel will not regulate independently, cross-talk between channels.
Root cause: Channel solenoid / transducer or a manifold fault.
Symptoms: Inflation-time alarm not sounding, timer wrong.
Root cause: Alarm / timer board fault - a safety function.
Symptoms: Leak at the connector, cuff will not hold.
Root cause: Worn tubing, connectors, or a damaged cuff.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Verify against a calibrated gauge after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Internal pump / compressor | Major service part | $$$ | Difficult |
| Solenoid / regulator valve | Major service part | $$ | Moderate |
| Pressure transducer Verify against a calibrated gauge after replacement. | Periodic replacement | $$ | Moderate |
| Connecting tubing / connectors | Frequent wear item | $ | Easy |
| Alarm / control board | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.