Datascope · Model Passport EL
ECG. Leads: I, II, III. Display sensitivity: 0.25, 0.5, 1.0, 2.0, 3.0, 4.0 cm/mV ± 10%. Frequency response to screen: ESU filter out: 0.5 to 40 Hz, -3db max @ 1 mVpp input; ESU filter in: 0.5 to 20 Hz -3db max @ 1 mVpp input Heart Rate Meter. Range: 30-300 bpm adult/child. 30-350 bpm neonatal. Accuracy: 30-250 bpm ± 3bpm or 3%, whichever is greater. Pacer rejection: rejects all pulse of amplitude ± 2.0 mV to ± 700 mV and duration 0.1 ms to 2 ms with no tail. Rejects all pulses ± 2.0 mV to ± 700 mV and duration 0.1 ms to 2 ms with 100 ms T.C. tail NIBP. Technique: oscillometric. Systolic range: adult/ped mode: 40-250 mmHg, neonate mode: 15-150 mmHg. Diastolic range: adult/ped mode: 30-200 mmHg, neonate mode: 10-150 mmHg. Systolic accuracy: mean error less than 7 mmHg, standard deviation less than 8 mmHg (adult/pediatric). Mean error less than 3 mmHg, standard deviation less than 13 mmHg (neonatal). Diastolic accuracy: mean error less than 7 mmHg, standard deviation less than 8 mmHg (adult/pediatric). Mean error less than 4 mmHg, standard deviation less than 7 mmHg (neonatal). Heart rate meter range: 30-250 bpm. Heart rate meter accuracy: ±3 bpm or ±3%, whichever is greater. Pulse Oximetry. Accuracy (1standard deviation): 70-100%, ±2% 60-69%, ±4% 0-59%, unspecified. Response time: 13 seconds to 95% of final step change % SpO2 value from 60-95% at 72 bpm.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: ECG, IBP, SpO2, temperature, CO2 or CO intermittently drops or is implausible.
Root cause: Worn patient cable / sensor, or a parameter-module / input-board fault.
Symptoms: NIBP times out, cuff will not inflate / hold, readings implausible.
Root cause: Internal NIBP pump or valve, a cuff / hose leak, or a pressure-transducer fault.
Symptoms: The host does not display a plugged-in module or a built-in parameter.
Root cause: Docking / connector fault or an internal board fault.
Symptoms: Blank, dim or distorted display; unresponsive touchscreen.
Root cause: Display, backlight, or controller fault.
Symptoms: Short battery run time, will not hold charge, will not power on.
Root cause: Aged battery pack or a power-supply fault.
Symptoms: Monitor not shown at the central station, admit fails, intermittent drop.
Root cause: Wired / Wi-Fi network configuration or a communication board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable set (ECG / IBP / SpO2 / temp) | Frequent wear item | $$ | Easy |
| NIBP cuff and hose | Frequent wear item | $ | Easy |
| Internal NIBP pump / valve | Major service part | $$$ | Difficult |
| Parameter module / input board | Major service part | $$$ | Difficult |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Display / touchscreen assembly | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.