Datascope · Model Passport 2 LT Monitor
Dedicated function keys provide one-touch access to all vital functions. Colorful waveforms with automatic display configuration simplify use and enhance display visibility. Standard features include six waveforms, auto-adjustable large numerics, 3 or 5-lead ECG, lead-selectable respiration, non-invasive blood pressure, temperature, and Masimo SET motion tolerant SpO2. Optional features include arrhythmia analysis, 3-lead ST analysis with trends and alarms, Microstream CO2, Nellcor OxiMax SpO2, two invasive blood pressures, and a dual trace recorder. View 12 ECG analysis module offers diagnostic quality 12-lead ECG interpretation with ST and arrhythmia analysis in a compact, easy-to-use module. Gas Module SE offers complete 5-agent auto-ID, CO2, O2 and N2O analysis for comprehensive and reliable gas monitoring. Built-in PCMCIA ports allow you to upgrade software, enhance memory, and transfer data with minimal effort. Integrated power supply with lithium-ion battery technology provides uninterrupted patient care during transport.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: ECG, IBP, SpO2, temperature, CO2 or CO intermittently drops or is implausible.
Root cause: Worn patient cable / sensor, or a parameter-module / input-board fault.
Symptoms: NIBP times out, cuff will not inflate / hold, readings implausible.
Root cause: Internal NIBP pump or valve, a cuff / hose leak, or a pressure-transducer fault.
Symptoms: The host does not display a plugged-in module or a built-in parameter.
Root cause: Docking / connector fault or an internal board fault.
Symptoms: Blank, dim or distorted display; unresponsive touchscreen.
Root cause: Display, backlight, or controller fault.
Symptoms: Short battery run time, will not hold charge, will not power on.
Root cause: Aged battery pack or a power-supply fault.
Symptoms: Monitor not shown at the central station, admit fails, intermittent drop.
Root cause: Wired / Wi-Fi network configuration or a communication board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable set (ECG / IBP / SpO2 / temp) | Frequent wear item | $$ | Easy |
| NIBP cuff and hose | Frequent wear item | $ | Easy |
| Internal NIBP pump / valve | Major service part | $$$ | Difficult |
| Parameter module / input board | Major service part | $$$ | Difficult |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Display / touchscreen assembly | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
One lot of "Low Level Output Cable - Interface to Philips Monitor" was received from the supplier with an incorrect cable in the packaging.
Non-functioning cable: the Low-Level Output Cable when connected to a Datascope Intra-aortic Ballon Pump (IABP) and patient monitor does not provide a signal to the patient monitor.
Software anomaly where the variable heart rates may be displayed inaccurately or intermittent 'dashes' may be displayed when the patient's heart rate is derived from ECG.
The Trio Monitor has latex feet. It is not labeled per 21 CFR801.437 to indicate that it contains latex.
Potential inaccurate printed information in the upper and lower margins of a print strip.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.