Unparalleled cardiac monitoring provides accurate data, helping you quickly make the best care decision. Legendary Marquette® ECG with EK-Pro™ and respirations offer comfort that lethal arrhythmias willbe brought to your attention. DINAMAP ASAP technology for accurate and reliable non-invasive BP monitoring. Configurations available with one of the following SpO2 solutions: Masimo® SET® (Gold Standard) or Nellcor OxiMax™. Configurations available with IVAC® TurboTemp™ for temperature in seconds.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: NIBP times out, cuff will not inflate / hold, readings implausible or fail against a reference.
Root cause: Internal NIBP pump or valve, a cuff / hose leak, or a pressure-transducer / calibration fault.
Symptoms: No saturation value or a poor pleth with a good site.
Root cause: Worn SpO2 sensor / cable, or the SpO2 module.
Symptoms: No or implausible temperature.
Root cause: Probe or cable fault.
Symptoms: Short battery run time, will not power on or charge.
Root cause: Aged battery pack.
Symptoms: No or faint strip, paper jam.
Root cause: Thermal print head or paper handling.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable set (ECG / IBP / SpO2 / temp) | Frequent wear item | $$ | Easy |
| NIBP cuff and hose | Frequent wear item | $ | Easy |
| Internal NIBP pump / valve | Major service part | $$$ | Difficult |
| SpO2 sensor / cable | Frequent wear item | $$ | Easy |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Thermal print head / paper | Periodic replacement | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio