Covidien · Model OptiVantage DH
Fully Programmable Powerhead Enables the technologist to stay with the patient longer. Powerhead features provide the ultimate in user-friendly operation. Saline can be programmed to inject prior to, or immediately following, contrast. Sleek and streamlined design for ergonomic convenience. All programming can also be achieved in the control room at the remote console. Perform a saline flush first to check for patent veins, and correct placement of the cannula, prior to contrast injection. Stay at patient’s side while testing. Perform a Timing Bolus injection of contrast medium followed by a saline flush to determine ideal scan delay for optimum image quality.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error message or code, injector will not arm or aborts an injection.
Root cause: Air in the syringe / line, a syringe not seated, or a subsystem fault indexed in the troubleshooting section.
Symptoms: Piston will not move, inaccurate flow or volume, stalls under pressure.
Root cause: Plunger drive motor, drive belt, or the plunger-position potentiometer.
Symptoms: No response, blank display, or arms but no delivery.
Root cause: Power supply, control board, or a head-to-console cable fault.
Symptoms: Contrast not warmed, heater indicator off or flashing.
Root cause: Failed heat maintainer or its cable.
Symptoms: Control panel text corrupt or the remote shows nothing.
Root cause: Display card or the communication path.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Plunger drive motor | Major service part | $$$ | Difficult |
| Plunger motor drive belt | Periodic replacement | $$ | Moderate |
| Plunger-position potentiometer | Major service part | $$ | Moderate |
| Syringe heat maintainer | Periodic replacement | $$ | Fairly easy |
| Disposable syringe set | Consumable | $ | Easy |
| Display / controller card | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio