MedRad · Model Spectris Solaris
3T compatibility - even when placed right next to the bore of the magnet Enhanced Power Management - through the integrated Continuous Battery Charger (iCBC) which allows you to plug the system into any available outlet Pressure Limit Control – preset pressure limits provide enhanced control for clinicians using additional patient access tubing like Power PICC and central lines Off Floor Mobile Design – easy to store flush against the wall.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: "CAL" system-malfunction code on the control panel, cannot arm.
Root cause: Calibration values missing or corrupt.
Symptoms: Control room unit and stand/head not communicating, no volume display, arm lights not sequencing.
Root cause: Fiber-optic cable fault or a misinstall, or a failed expansion-base / stand-base interface card or Servo/CPU card.
Symptoms: Battery indicator not filled on the CRU screen, head loses power, iCBC not charging.
Root cause: Aged SRU battery or a charger fault.
Symptoms: Erratic or stalled plunger travel, volume values not changing when plungers are moved manually.
Root cause: Encoder or power-drive card fault on motor A/B.
Symptoms: Under-volume, air-column warnings, leaks at the patient tubing.
Root cause: Improperly loaded syringe or a worn/cracked disposable connector set.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibrate after fitting.
Single-patient consumable.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Fiber-optic cable (CRU-SRU) | Major service part | $$ | Moderate |
| Servo / CPU card Recalibrate after fitting. | Major service part | $$$$ | Difficult |
| Expansion-base / stand-base interface card | Major service part | $$$ | Difficult |
| SRU battery | Periodic replacement | $$ | Fairly easy |
| MR-compatible disposable syringe and connector tubing Single-patient consumable. | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.