BRACCO · Model Empower MR
ARM mode allows software to run diagnostic checks prior to injection, as well as guarding against unintentionally starting the injection Tilt lockout minimizes the risk of air embolism by ensuring that the injector runs only in the down position Color-coded illuminating handle lets you know whether the injector Random Access Memory (RAM) is in motionUser-friendly remote display with anatomical identifiers allows easy viewing and programming of protocols with up to 8 phases (3 phases when saline is included) Stores up to 50 separate programs Easily accessible Help function provides instant on-screen operational guidance.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: ARM mode diagnostic checks fail, injector will not go to armed.
Root cause: Air detected in the syringe/line, a syringe not seated, or a sensor/self-test fault.
Symptoms: Injector will not run, tilt-lockout indication.
Root cause: Injector head not in the down position - the tilt sensor prevents running to reduce air-embolism risk.
Symptoms: Plunger will not move, inaccurate volume, stalls.
Root cause: Drive motor, leadscrew, or plunger-engagement fault.
Symptoms: No status on the remote, illuminating handle does not indicate motion, cannot program.
Root cause: Head-to-remote cable or a communication board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Syringe drive motor / leadscrew | Major service part | $$$ | Difficult |
| Disposable syringe set | Consumable | $ | Easy |
| Hand / pendant control | Periodic replacement | $$ | Fairly easy |
| Tilt sensor | Major service part | $$ | Moderate |
| Head-to-remote cable | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Synced live from the FDA openFDA API.
Insufficient seal on the sterile barrier of the device, compromising the sterility of the device
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio