MedRad · Model Avanta
Ease-of-use On-screen tutorial for simplified setu Single-patient and multi-patient disposables (up to five patients Pedestal or table mounting with independent mounting of display and injection uni Easy-to-read, color touchscreen\ Safety Precise control of coronary injection, especially with small catheters Accurate injection pressure control with user-adjustable pressure limits Redundant air management system with level sensing and gross air detection Reduced radiation exposure for operator Precision Distinct bolus sharpness feature Exact variable and fixed-flow delivery Increased accuracy and responsiveness with hand controller (when compared to hand injection) Automatic transition from injection to hemodynamic state.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Injector will not advance the plunger, stalls, pressure-limit or motor-fault alarm.
Root cause: Drive-motor / lead-screw wear, a syringe-interface / plunger-ram fault, or a pressure-transducer fault.
Symptoms: Air-detector alarms with no air, or fails to detect air, will not arm.
Root cause: Air-detector sensor or a dirty / damaged sensor window.
Symptoms: The console loses the injector head, hand-controller unresponsive, arm / pause not working.
Root cause: Head-to-console cable / wireless link or a board fault.
Symptoms: Contrast not warmed, warmer alarm.
Root cause: Syringe / bath heater or its sensor.
Symptoms: Short run time, will not hold charge.
Root cause: Aged battery pack.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Drive motor / lead-screw assembly | Major service part | $$$ | Difficult |
| Plunger ram / syringe-interface assembly | Major service part | $$ | Moderate |
| Pressure transducer | Major service part | $$ | Moderate |
| Air-detector sensor | Major service part | $$ | Moderate |
| Battery pack (mobile heads) | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Synced live from the FDA openFDA API.
The seal between the white polyethylene and the clear plastic may be missing on the Hand Controller Sheath package, resulting in a potential breach of the sterility of this accessory.
Event was initially intended to be a product enhancement of the MEDRAD Avanta Fluid Management Injection System. Upgrades were initiated in October, 2007. Observance of a possible increase in complaints and adverse event reporting resulted in the firm issuing "Important Product Use Reminder" to remind users of proper procedures for using the system and the need for diligence in reducing air embo
The product may be defective resulting in a reduction of the saline delivery rate and inadequate air purging.
Sterility (package integrity) compromised: Incomplete package seal; sterility may be compromised
stress crack on the stopcock
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.