BRACCO · Model EmpowerCTA+
Unique Safety Enhancements: Saline advance - Delivered via pendant switch or key on the injector at the patient's side prior to starting the procedure, allows for the use of saline to test patency and vein integrity Saline jump - Allows for the immediate advancement to saline (once opacification has been reached), minimizing the contrast a patient receives Unique patented EDATM (extravasation detection accessory) Detects, notifies, and pauses procedure if clinically significant variation occurs; detects extravasation with patented EDA patch, providing additional patient safety and peace of mind for the technologist eGFR calculator - Assists in the critical evaluation of kidney function before contrast is delivered Streamlined Workflow: Simultaneously - operating dual syringes Auto-initialization - The most efficient way to prepare syringes for filling Protocol fill - Automatically fills syringes to volumes based on the selected protocol Integrated Efficiencies: Built-in tilt sensor and position lock to minimize the likelihood of an air embolism Convenient remote control pause function Fixed fill rate Purge function to automatically remove air from the system.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Pre-injection checks fail, will not arm.
Root cause: Air in a syringe or line, a syringe not seated, or a sensor/self-test fault.
Symptoms: Injector will not run.
Root cause: Head not in the down position - the built-in tilt sensor and position lock guard against air embolism.
Symptoms: EDA accessory not recognised, false extravasation pauses, or no detection.
Root cause: EDA patch not connected or defective, or the EDA input.
Symptoms: One syringe will not fill or advance, volume inaccurate, purge fails.
Root cause: Drive motor, leadscrew, or plunger engagement on that channel.
Symptoms: No remote control, pause function unresponsive.
Root cause: Cable or communication board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Syringe drive motor / leadscrew (per channel) | Major service part | $$$ | Difficult |
| Disposable dual-syringe set | Consumable | $ | Easy |
| EDA patch / accessory | Consumable | $ | Easy |
| Hand / pendant control | Periodic replacement | $$ | Fairly easy |
| Tilt sensor / position lock | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Synced live from the FDA openFDA API.
Insufficient seal on the sterile barrier of the device, compromising the sterility of the device
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio