Cecil Instruments · Model Super Aurius
Calibration Curves Multi wavelength measurements DNA and protein assays Cell and wavelength programs Kinetics and band areas.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Low-energy warning, noisy baseline, poor signal-to-noise, especially in the UV.
Root cause: Deuterium (UV) or tungsten-halogen (visible) lamp at end of life.
Symptoms: Baseline will not flatten, absorbance offset, peaks shift.
Root cause: Contaminated sample-compartment optics or a dirty cell holder.
Symptoms: Holmium or known-peak check fails.
Root cause: Monochromator wavelength-drive calibration.
Symptoms: Software will not connect or control the instrument.
Root cause: Interface cable, port, or driver.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Deuterium (UV) lamp | Frequent wear item | $$$ | Moderate |
| Tungsten-halogen (visible) lamp | Periodic replacement | $$ | Moderate |
| Sample cell holder | Major service part | $ | Easy |
| Communication cable | Periodic replacement | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.