Bio-Rad Laboratories, Inc. · Model xMark
Monochromator-based system for maximum optical efficiency and flexibility in photometry reading Spectral scanning to define the best absorbance wavelength for assays Imaging capabilities for pixel intensity data in a single well or entire plate Ability to scan custom plates Flexible control and data analysis with PC or Mac software Incubator with programmable temperature control Compatible with 6- to 1,536-well and custom microplates 128 mm L x 88 mm W Plate height cannot exceed 21 mm Clear flat-bottom plates are recommended such as Bio-Rad's Costar 96–Well Flat–Bottom EIA Plate (2240096) U-Bottom plates can be used, but V-bottom plates are not recommended.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Blank or standard absorbance off, poor plate-to-plate reproducibility, high CV.
Root cause: Aged lamp / light source, monochromator drift, contamination on the optics, or a dirty plate transport.
Symptoms: Plate not loading, jams, mis-registration of wells, "plate error".
Root cause: Worn transport belt/motor or a position-sensor fault; a plate taller than 21 mm or a V-bottom plate used.
Symptoms: Chamber not reaching or holding the programmed temperature, assay results temperature-sensitive.
Root cause: Heater element, temperature sensor, or control-board fault.
Symptoms: Software will not connect, control or data transfer fails.
Root cause: USB/serial cable or port fault, or a driver issue.
Symptoms: Spectral scan noisy or shifted, wrong absorbance wavelength selected.
Root cause: Monochromator grating drive or its calibration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Lamp / light source | Periodic replacement | $$ | Moderate |
| Plate transport belt / motor | Periodic replacement | $$ | Difficult |
| Incubator heater / temperature sensor | Major service part | $$ | Moderate |
| Plate position sensor | Periodic replacement | $ | Moderate |
| Communication cable | Periodic replacement | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.