Carestream · Model DRX Evolution Plus
The new DRX-Evolution Plus offers New LED lighting for enhanced functionality and aesthetics Greater flexibility with an extended tube column High performance Carestream generator Optional table to accommodate patients up to 705 lbs (320 kg) Forward-looking design to accommodate advanced imaging applications in the future Proven Image Quality The wireless, cassette-sized DRX Detector and EVP Plus image-processing software deliver superb image quality. Optional EVP Plus Software provides increased latitude and high-contrast image detail. Software options accommodate specialties such as pediatrics, orthopedics, and trauma. Versatility Choose a one-, two- or three-detector option to perform many types of imaging exams, including tabletop and long-length imaging. The automated overhead tube is designed for positioning convenience and efficiency. Detectors are fixed or wireless. Scalability Customize a DR solution to fits your workflow and budget. Add DR systems as your facility expands and grows. Designed for advanced future applications. Increased Productivity Images are ready for viewing and manipulation on the console and in seconds. One space-saving console controls all medical imaging functions. Auto-positioning for fast and accurate setup.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Panel not detected, images not received, wireless / tethered link drops mid-exam.
Root cause: Detector battery (wireless), Wi-Fi / tether cable, a docking / connector fault, or detector electronics.
Symptoms: Fixed-pattern artefacts, non-uniformity, dead pixels / lines, calibration overdue or failing.
Root cause: Detector gain / offset calibration, a bad pixel map, or physical detector (drop) damage.
Symptoms: Exposure inhibited, kV / mAs error, no exposure captured, AEC over/under.
Root cause: HV generator, tube, filament supply, or an AEC / interface fault.
Symptoms: Tube column, wall stand or table bucky will not track, detent, or auto-position.
Root cause: Drive motor, brake, encoder, or the auto-positioning interface.
Symptoms: Images will not display, process or send; worklist missing; slow processing.
Root cause: Workstation PC, software, or the DICOM / network configuration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Flat-panel detector (exchange) | Major service part | $$$$ | Moderate |
| Detector battery (wireless) | Periodic replacement | $$ | Easy |
| Tether cable / access point | Periodic replacement | $$ | Fairly easy |
| Positioner drive / brake / encoder | Major service part | $$$ | Difficult |
| Acquisition workstation PC | Major service part | $$$ | Moderate |
| AEC ion chamber | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio