BTL · Model 5000
High pressure up to 5 bars. High frequency up to 22 Hz. Burst mode for extra-sensitive patients. 8.4″ colour touch screen. Colour therapeutic encyclopaedia with anatomical images. User-defined protocols. Modular system – possibility of later upgrade with electrotherapy, ultrasound, laser, or magnetotherapy. Supplied with the most comfortable applicator on the market – special ergonomic grip and elimination of backward shocks.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Ultrasound, electrotherapy, laser or shockwave module delivers no or weak output, error on that channel.
Root cause: Applicator / handpiece / electrode fault, a cable fault, or an output-stage fault on that module.
Symptoms: Low or no ultrasound output, applicator-contact / BNR error, no warmth on a test.
Root cause: Worn or cracked ultrasound applicator crystal, or a cable fault.
Symptoms: No stimulation on a channel, output will not ramp, patient feels nothing.
Root cause: Lead wires, dried electrodes, an open circuit, or an output-stage fault.
Symptoms: Unresponsive or erratic touchscreen, parameters will not set.
Root cause: Touchscreen or controller fault.
Symptoms: Unit will not power on, or the powered-height trolley will not adjust.
Root cause: Power supply / fuse, or the trolley actuator.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Ultrasound applicator | Periodic replacement | $$ | Easy |
| Electrotherapy lead wires / electrodes | Frequent wear item | $ | Easy |
| Laser / shockwave handpiece | Periodic replacement | $$$ | Fairly easy |
| Touchscreen assembly | Major service part | $$$ | Moderate |
| Power supply / fuse | Major service part | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.