BPL Medical Technologies · Model ExcelSign E17
Modular platform well suited for high acuity and operating room environment. ExcelSign E17: 17″ High Resolution Touchscreen Display. Standard Module: Ensures increased clinical confidence in measuring the vital parameters - ECG, RESP, SpO2, NiBP, 2-IBP, TEMP. Transport Monitor (Optional): Equipped with Display, independent operating system with in-built battery. Enhanced Connectivity - USB patient data output, supports external printer, HL7, FTP server upload, VGA for external display, optional Medius+ central nursing station. Efficient memory storage with Calculation Interface. High performance Li-ion battery.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Flat or noisy ECG, frequent lead-off alarms.
Root cause: Worn electrodes, loose leads, or dry gel.
Symptoms: No pleth, SpO2 drops to --- or false low.
Root cause: LED/photodiode aging or cable damage.
Symptoms: Cuff fails to inflate, erroneous BP, long cycle.
Root cause: Pump/valve wear or cuff leak.
Symptoms: Short backup time, battery alarm on transport.
Root cause: Aged backup battery.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable per patient.
High-turnover; keep inventory.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| NIBP cuff & hose Consumable per patient. | Periodic replacement | $ | Easy |
| Rechargeable backup battery | Periodic replacement | $$ | Fairly easy |
| Display / touch assembly | Periodic replacement | $$$ | Difficult |
| ECG patient cable & lead set High-turnover; keep inventory. | Frequent wear item | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.