BPL Medical Technologies · Model Excello
10.4-inch High resolution display. Standard Configuration - ECG, RESP, SPO2 (Masimo / Nellcor / Digital), NiBP, TEMP. Optional Parameters - Dual IBP, ETCO2, AGM, ICG, Thermal Recorder. High resolution 10.4” colour TFT display Standard configuration of ECG, RESP, Nellcor SpO2 / Masimo SpO2/ Digital SpO2, NiBP, TEMP The added functions of the device are drug dose calculation / HRV analysis / OXY-CRG The device provides for arrhythmia detection and ST analysis Lead ECG / 3 Lead ECG (Optional) Maximum 10 waveforms display Provides for maximum of 168 hours of trend review Detects pacemaker pulses >120 minutes operating time in fully charged condition >240 minutes with additional battery (optional) Wired or Wireless networking with Central monitoringsystem software (optional) 2-IBP, ETCO2( mainstream/side stream/ micro stream), AGM, ICG and thermal recorder (optional)
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Flat or noisy ECG, frequent lead-off alarms.
Root cause: Worn electrodes, loose leads, or dry gel.
Symptoms: No pleth, SpO2 drops to --- or false low.
Root cause: LED/photodiode aging or cable damage.
Symptoms: Short backup time, battery alarm on transport.
Root cause: Aged backup battery.
Symptoms: Dim, flickering, or unresponsive screen.
Root cause: Backlight or digitizer failure.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable per patient.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| SpO2 sensor (adult/ped/neo) | Frequent wear item | $$ | Easy |
| NIBP cuff & hose Consumable per patient. | Periodic replacement | $ | Easy |
| Rechargeable backup battery | Periodic replacement | $$ | Fairly easy |
| Display / touch assembly | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.