Biegler · Model BW 485L
Temperature can be set from 37 degrees C to 41 degrees C in 0,5 degree C increments Horizontal heating cylinder facilitates operation Several transfusions/infusions can be warmed at the same time Easy to read display Splash waterproof (IPX4) Low temperature alarm Extremely safe with permanently running self-tests Integral service interface for PC
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: No heat, or overheat alarm.
Root cause: Heater element or thermal cutoff failure.
Symptoms: Delivered fluid temperature out of set range.
Root cause: Inline/probe sensor aging or miscalibration.
Symptoms: False or missed air-in-line alerts.
Root cause: Air detector sensor drift or dirty optics.
Symptoms: Fluid leak, or won't prime.
Root cause: Worn cassette/tubing seat or cracked fitting.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Single-use consumable.
Consumable in some designs.
Safety-critical; verify at PM.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Disposable tubing / cassette set Single-use consumable. | Consumable | $ | Easy |
| Heating element / plate | Periodic replacement | $$$ | Moderate |
| Temperature sensor / probe Consumable in some designs. | Periodic replacement | $$ | Fairly easy |
| Air-in-line sensor Safety-critical; verify at PM. | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.