Arjo · Model AtmosAir 9000
Non-powered system Sloped heel section Firm perimeter 9 horizontal air cylinders Pressurized system comprised of independent, dynamic air chambers and intake and regulator valves.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Surface stays soft, patient bottoms out, will not firm up.
Root cause: Weight not opening the Self-Adjusting Technology (SAT) valves, internal tubing kinked or disconnected, or holes/damage in the SAT system.
Symptoms: Splits or cracking in the waterproof cover, staining, moisture inside.
Root cause: Wear or sharp objects - the mattress must not be used with a compromised cover.
Symptoms: One zone flattens, uneven support.
Root cause: Perforated or split air cylinder / bladder.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Mattress cover | Periodic replacement | $$ | Fairly easy |
| SAT air cylinder / bladder set | Periodic replacement | $$$ | Moderate |
| Internal connection tubing / connectors | Major service part | $ | Moderate |
| Intake / regulator valve | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.