Amico · Model Powered Therapeutic Mattress
Vented cushions for low air loss therapy preventing moisture build-up Microclimate management increases air flow and circulation 16 cell overlay allows for even distribution of patient’s weight and maximizes pressure relief Anatomical uninterrupted pressure distribution Easy to set up and clean Lightweight design with handles allows for effortless transport and storage Lock-out key prevents tampering of patient settings and power switch Accommodates patient weight of up to 500 lbs (226.8 kg)
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Surface stays flat, pump runs but no pressure, or pump does not run.
Root cause: Blower/pump fault, a hose disconnect or kink, or a power fault.
Symptoms: Localised loss of support, uneven surface.
Root cause: Perforated air cell or a failed cell connector.
Symptoms: Patient feels damp, little airflow through the vented cushions.
Root cause: Air filter clogged or blower output low.
Symptoms: Surface does not deflate quickly when the CPR control is used.
Root cause: CPR dump valve or the connecting hose.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Pump / blower unit | Periodic replacement | $$$ | Moderate |
| Air cell / bladder (16-cell overlay) | Periodic replacement | $$ | Fairly easy |
| Hose and connectors | Periodic replacement | $ | Fairly easy |
| Air intake filter | Frequent wear item | $ | Easy |
| Mattress cover | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.