Hillrom · Model Envision E700
Advanced Microclimate Technology Uses independent air cells to immerse and envelop for best in class pressure redistribution ShearRelief® System minimizes the effects of shear and friction during patient movement and bed articulation.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Patient bottoms out, low-pressure alarm, surface stays soft.
Root cause: Blower / pump fault, a hose disconnect or kink, or an air-cell / bladder leak.
Symptoms: Localised loss of support.
Root cause: Perforated air cell or a failed cell connector.
Symptoms: Patient feels damp, little airflow through the coverlet.
Root cause: Clogged air filter or low blower output.
Symptoms: Surface does not deflate quickly on the CPR control.
Root cause: CPR dump valve or hose fault.
Symptoms: Tears or cracking in the top cover, fluid ingress.
Root cause: Wear or sharp objects.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Pump / blower unit | Periodic replacement | $$$ | Moderate |
| Air cell / bladder | Periodic replacement | $$ | Fairly easy |
| Hose and connectors | Periodic replacement | $ | Fairly easy |
| Air filter | Frequent wear item | $ | Easy |
| Mattress top cover | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio