
American Diagnostic Corporation · Model Adstation 5660TE
Adstation Wall Transformer with one extension module features: Magnetic switch activates power when handle is removed from cradle, automatically powers off when handle is replaced Handle CNC machined from solid brass, the handle is chrome plated in black with a ribbed surface for a sure grip Adtronic electronic rheostatic light intensity control provides infinite continuous control Adtronic auto-off feature after 2 minutes provides additional safety if instrument is mis-cradled Bayonet instrument mount is compatible with all major brands of 3.5v instrument heads Base unit contains power switch and one handle Up to 2 additional extension modules (3 total) can be added Easy to install with included bracket or may be mounted to our wallboard systems Available in 120v or 230v power supply with Euro plug UL approved Five year warranty.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Otoscope / ophthalmoscope light stays full or off, will not dim from the handle rheostat.
Root cause: Worn handle rheostat, a bulb at end of life, or a wall-transformer output fault.
Symptoms: Light flickers or cuts out when the spiral cord is stretched.
Root cause: Broken conductors in the coiled cord.
Symptoms: BP gauge reads high or low vs a reference.
Root cause: Wear or shock to the aneroid movement.
Symptoms: No power to the instrument heads.
Root cause: Failed wall transformer or its wiring.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Instrument-head bulbs | Periodic replacement | $ | Easy |
| Handle rheostat | Major service part | $$ | Moderate |
| Coiled handle cord | Periodic replacement | $$ | Fairly easy |
| Wall transformer | Major service part | $$ | Moderate |
| Aneroid BP gauge / cuff (T/TE models) | Major service part | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio