AGFA · Model CR 30-X
High image quality and throughput -The CR 30-X scans at a spatial resolution of 10 pixel/mm, then processes the image using MUSICA² - Agfa HealthCare's patented, market-leading intelligent image processing software. By scanning the phosphor plate at the full width of 43 cm, the CR 30-X provides both high quality and high throughput. The CR 30-X supports all common cassette formats. Low total cost of ownership - Installing the CR 30-X digitizer can be achieved in a single day. Its special LED technology in the erasure unit keeps the heat output and the power requirements lower, so that standard electrical outlets are sufficient. The lack of preliminary electrical work means the set-up cost is lower. And with its modular, component-based design, the CR 30-X digitizer offers faster, easier and more cost-effective maintenance. Facilitates convenient and fast workflow- The CR 30-X works in conjunction with Agfa HealthCare's NX - the CR's image identification and quality control workstation designed to enable users to complete effortlessly, and with minimal training, a complete digital X-ray exposure. In addition, the CR 30-X uses MUSICA intelligent image processing software to automatically optimize the final image quality without the need for human intervention. Designed to enable a seamlessly integrated digital workflow, the CR 30-X uses DICOM protocols for all steps in the digital workflow.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Banding, streaks, ghost images, non-uniform density, dust artefacts.
Root cause: Dirty light guide / scanning optics, a worn erase-lamp array, imaging-plate wear, or transport-roller contamination.
Symptoms: Plate jams in the reader, transport error, plate not returned.
Root cause: Worn or contaminated transport rollers / belts, or a sensor fault.
Symptoms: Faint previous image on the next exposure.
Root cause: Erase-lamp array degradation or a driver fault.
Symptoms: No or very noisy image, scan-error message.
Root cause: Scanning laser or the photomultiplier / detection chain.
Symptoms: Images will not display, annotate or send to PACS.
Root cause: Workstation PC or the DICOM configuration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Transport rollers / belts | Periodic replacement | $$ | Moderate |
| Erase-lamp array | Periodic replacement | $$$ | Moderate |
| Imaging plates / cassettes | Periodic replacement | $$$ | Easy |
| Light guide / scanning optics | Major service part | $$$ | Difficult |
| Workstation PC | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.