3M Health Care · Model V.A.C.ULTA"
Cleanse the wound with the instillation of topical wound cleansers Treat the wound with the instillation of topical antimicrobial and antiseptic wound solutions Heal the wound and prepare for closure Contaminated wounds with instillation of topical wound cleansers and Infected wounds with instillation of appropriate topical wound solutions and in conjunction with appropriate good clinical practice such as debridement and antibiotic therapy Volumetric delivery Fill Assist to help determine the correct instill volume Dressing Soak to instill topical wound solutions into the wound for easier dressing removal and increased patient comfort.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Low-pressure / leak alarm, therapy pauses, dressing not collapsing.
Root cause: Dressing / canister leak, worn tubing / connectors, a worn pump, or a pressure-sensor fault.
Symptoms: "Canister full" or "blockage" alarm, therapy stops.
Root cause: Full or mis-seated canister, a clot / gel in the tubing, or a canister-sensor fault.
Symptoms: Short run time, will not reach full charge.
Root cause: Aged battery pack.
Symptoms: Cannot build pressure with a sealed test, pump noisy or dead.
Root cause: Worn pump / diaphragm or motor.
Symptoms: Unresponsive keys, blank display, no audible alarm.
Root cause: Display / keypad, or the alarm / control board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Pump / diaphragm assembly | Periodic replacement | $$ | Moderate |
| Pressure sensor | Major service part | $$ | Moderate |
| Tubing / connector set | Frequent wear item | $ | Easy |
| Display / control board | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.