
X-CEL · Model X-Cel X-Ray MB 700 A/C
Eliminate the need to rotate the patient for the DP view Eliminate the need to rotate the arm for the AP view Eliminate the need to readjust the collimator lateral views Enable wheelchair patients to be easily accommodated without having to leave the wheelchair Feature the low voltage expose switch with optional remote key box Include features that can meet ADA guidelines 28-inch SID Saves floor space 3.5-inch base height Fits between wall and chair Minimum space required: 34″ x 51″ Both laterals possible Folds to 18 inches wide 15 mA 50-70 kVp.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Exposure inhibited, generator error / fault code, images consistently under- or over-exposed.
Root cause: HV generator / inverter fault, an exposure-switch / contactor fault, an HT-cable fault, or a tube-load / thermal interlock.
Symptoms: Output does not match the console, techniques drift, generator calibration fails.
Root cause: Inverter / HV-tank fault, a filament-supply fault, or a calibration issue.
Symptoms: Auto-exposure images consistently over- or under-exposed, chambers not tracking.
Root cause: AEC ion chamber(s) or the AEC calibration.
Symptoms: Column, wall stand, table or bucky will not track / detent; collimator light out; grid errors.
Root cause: Drive / brake / detent hardware, the collimator lamp / drive, or a bucky-motor fault.
Symptoms: Exposures inhibited after a series, tube-heat alarm.
Root cause: Tube housing heat limit reached or a cooling-fan fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| HV generator / inverter components | Major service part | $$$$ | Expert-level |
| HT cables | Major service part | $$$ | Difficult |
| X-ray tube assembly | Major service part | $$$$ | Expert-level |
| AEC ion chamber | Major service part | $$$ | Difficult |
| Collimator lamp / drive assembly | Periodic replacement | $$ | Moderate |
| Bucky / grid motor | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
The X-Ray system contained a defective X-Ray tube shield.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio