Shimadzu · Model RADspeed
Effortless tube positioning allows the operator to focus on patient care. Manual operation is also possible to make corrections of the fine-tuning positioning extremely simple Ensures high-quality radiographic images at the optimal X-ray dose. Enables fast positioning at complex angles for orthopedic applications. Up to 800 anatomical programs can be registered on the system The LCD screen and illumination colour can change according to the Bucky table or X-ray tube settings selected Different alarm sounds can also be specified for various events, e.g. when preparation for exposure is complete. Combination with DR system optimises your workflow Images are taken while swinging the X-ray tube and moving the FPD. These images are automatically stitched together to obtain wide-range images along the body axis instantly.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Image noise, artifacts, low SNR.
Root cause: Flat-panel / sensor aging.
Symptoms: Tube temp alarm, duty-cycle limited.
Root cause: Pump/radiator or coolant loss.
Symptoms: Exposure fails, kV/mA out of range error.
Root cause: HV tank or generator board fault.
Symptoms: Frozen console, acquisition timeout.
Root cause: Workstation or RAID degradation.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Major R&R; track mAs history.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Workstation RAID / GPU | Periodic replacement | $$$ | Difficult |
| X-ray tube insert Major R&R; track mAs history. | Major service part | $$$$ | Expert-level |
| Flat-panel detector | Major service part | $$$$ | Expert-level |
| Cooling pump & radiator | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →