WEM · Model SE-120
Indispensable accessory for any kind of surgical procedures. Continuous operation equipment, connected by means of an electric switch. Suction is by means of a vacuum-generating fan, driven by a 127 VAC or 220 VAC electric motor with automatic voltage selection. The apparatus is positioned and installed next to the electrosurgical unit and, in order to provide suction to the required site; accessories such as a hose and a tapered nozzle or a speculum with tube are provided. A high-efficiency (HEPA) mechanical filter is placed at the unit's suction inlet, upstream from the accessories, to trap aspirated viruses and bacteria. An additional activated-charcoal chemical filter is fitted to eliminate the odors from cutting and coagulation with the electrosurgical unit.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Plume not captured, weak airflow, blower ramps but flow is low.
Root cause: Loaded ULPA / charcoal filter, a worn blower / turbine, or a leak / kink in the hose.
Symptoms: Filter-change alarm, reduced airflow, odour breakthrough.
Root cause: ULPA / activated-charcoal filter at end of its capacity.
Symptoms: Evacuator does not start with the ESU, or runs continuously.
Root cause: The ESU-sensing cuff / cable or the auto-activation board.
Symptoms: Noisy, weak, or dead blower even with a new filter.
Root cause: Worn blower / turbine bearings or motor.
Symptoms: Very loud in use, speed will not adjust.
Root cause: Worn blower or a speed-control fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| ULPA / activated-charcoal filter | Consumable | $$ | Easy |
| Blower / turbine assembly | Periodic replacement | $$$ | Difficult |
| ESU-sensing cuff / auto-activation board | Major service part | $$ | Moderate |
| Hose / wand adapters | Frequent wear item | $ | Easy |
| Speed-control board | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio