Conmed · Model 1000 SES
The first stage filtration is a prefilter whose function is to trap and remove gross particulate and casual fluid. The second stage filtration is a VLSI (Very Large Scale Integrated) grade ULPA (Ultra Low Penetration Air) filter whose high-tech patented design captures particulates and micro-organisms down to 0.012 µm at an efficiency of 99.99995%. Each VLSI grade ULPA filter is pretested and certified before use. The third stage filtration uses the highest grade virgin coal grade activated carbon for the removal and adsorption of odors and toxic gases produced by cauterization of tissue. These harmful gases may constitute a health hazard to healthcare professionals who are subjected to prolonged exposure. The activated carbon used in the CONMED 1000SESTM preferentially removes toxic organic gases rather than water vapor and provides optimal odor removal. The fourth stage filtration is an expanded foam and paper media used to trap activated carbon fines from migrating out of the filter. The first, second and fourth stages of the CONMED 1000SESTM Filter have been impregnated with a chemical agent. The chemical agent has been applied with a patented process that prevents the agent from being washed or blown off the filter media.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Plume not captured, weak airflow, blower ramps but flow is low.
Root cause: Loaded ULPA / charcoal filter, a worn blower / turbine, or a leak / kink in the hose.
Symptoms: Filter-change alarm, reduced airflow, odour breakthrough.
Root cause: ULPA / activated-charcoal filter at end of its capacity.
Symptoms: Evacuator does not start with the ESU, or runs continuously.
Root cause: The ESU-sensing cuff / cable or the auto-activation board.
Symptoms: Noisy, weak, or dead blower even with a new filter.
Root cause: Worn blower / turbine bearings or motor.
Symptoms: Very loud in use, speed will not adjust.
Root cause: Worn blower or a speed-control fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| ULPA / activated-charcoal filter | Consumable | $$ | Easy |
| Blower / turbine assembly | Periodic replacement | $$$ | Difficult |
| ESU-sensing cuff / auto-activation board | Major service part | $$ | Moderate |
| Hose / wand adapters | Frequent wear item | $ | Easy |
| Speed-control board | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.