Surveyor Central Station with S4 Wifi telemetry and S12/S19 multiparameter bedside monitoring capabilities; English language and CD manual; Configured CPU (custom SCNODE configuration must be determined) with 110V power and dual color LCD 24" monitors with high resolution; 0 hard wired patient monitoring slots and 4 telemetry channels (single display required); Local storage for up to 600 days, 110V HP LaserJet network printer; VERITAS analysis algorithm with 12-lead ECG interpretation, extende
Intended use: A mains electricity (AC-powered) device designed to receive continuous, interval and/or spot-check physiological data (e.g., vital signs) from one or more bedside patient monitor(s) (not included) to display this information at a central patient monitoring station where staff can monitor multiple patients simultaneously. It is comprised of hardware with integrated software and is especially used in critical care settings, producing visible and/or audible signals/alarms when adverse conditions are registered. Additional features may include Holter/ST-segment monitoring and patient data transfer [e.g., into an electronic medical record (EMR)].
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Flat or noisy ECG, frequent lead-off alarms.
Root cause: Worn electrodes, loose leads, or dry gel.
Symptoms: No pleth, SpO2 drops to --- or false low.
Root cause: LED/photodiode aging or cable damage.
Symptoms: Cuff fails to inflate, erroneous BP, long cycle.
Root cause: Pump/valve wear or cuff leak.
Symptoms: Dim, flickering, or unresponsive screen.
Root cause: Backlight or digitizer failure.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
High-turnover; keep inventory.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Rechargeable backup battery | Periodic replacement | $$ | Fairly easy |
| Display / touch assembly | Periodic replacement | $$$ | Difficult |
| ECG patient cable & lead set High-turnover; keep inventory. | Frequent wear item | $$ | Easy |
| SpO2 sensor (adult/ped/neo) | Frequent wear item | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.