A-DEC, INC. · Model 8800B
Performer 8000 Monitor Mount / 8800B.
Intended use: A permanently-fixed structure designed to support a light and/or an image display monitor that is used for medical purposes [e.g., in an operating room (OR) and/or an examination/treatment room]. The mount is intended to be permanently-fixed to the ceiling and/or wall. This device will be designed to enable dynamic positioning (e.g., with dynamic arms, sliding bracket); the device to be mounted (medical light/monitor) is not included.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Base, back or a positioning function does not respond to the foot control or touchpad.
Root cause: Foot control / touchpad fault, a hydraulic valve or actuator, or a control-board fault.
Symptoms: Base slowly lowers under load after positioning.
Root cause: Internal leak in the hydraulic cantilever, a failing check valve, or the stop-plate shut-off.
Symptoms: Programmed positions do not recall or the chair stops partway.
Root cause: Program board or a position-feedback fault.
Symptoms: Tilting back or the articulating headrest slips under load.
Root cause: Worn lock or friction hardware.
Symptoms: Loud pump on lift, slow rise.
Root cause: Worn pump or low hydraulic fluid.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Hydraulic pump / motor | Major service part | $$$ | Difficult |
| Hydraulic control valve | Major service part | $$ | Difficult |
| Foot control / programmable footswitch | Periodic replacement | $$ | Fairly easy |
| Back / tilt actuator | Major service part | $$$ | Difficult |
| Headrest lock hardware | Major service part | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.