Philips Healthcare · Model MX450
Compact multi-measurement vital signs monitor with touchscreen, part of the IntelliVue family, supports ECG, SpO2, NIBP, temperature, and more.
Intended use: Continuous monitoring of adult/pediatric/neonatal vital signs in hospital environments.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: '--' SpO2, frequent low-signal alarms
Root cause: Worn connector pins or damaged extension cable
Symptoms: NIBP measurement failures, over-pressure errors
Root cause: Cracked tubing, worn cuff bladder
Symptoms: Noisy ECG trace, false asystole alarms
Root cause: Fatigued shielded leadwires / dried electrodes
Symptoms: Runtime < 30 min, battery warning icon
Root cause: Li-ion aging past ~500 cycles
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Rechargeable Li-ion battery pack | Periodic replacement | $150–$300 | Easy |
| NIBP hose assembly | Periodic replacement | $60–$120 | Easy |
| SpO2 extension cable | Frequent wear item | $80–$160 | Easy |
| ECG trunk cable (5-lead) | Frequent wear item | $90–$200 | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →