WELCH ALLYN, INC. · Model 901060
Welch Allyn Connex 6800 Vital Signs Monitor with Masimo SpO2, Masimo Acoustic Respiratory Monitoring (RRa), SureTemp Plus Thermometry, SureBP Non-invasive Blood Pressure, Pulse Rate, MAP, BP Averaging, Nurse Call, Four USB Ports for Accessories, Internal 802.11 a/b/g Wireless Radio; 100-240 V, 50-60 Hz AC, IEC Plug Type-B
Intended use: An electrically-powered device/device assembly designed for the continuous assessment of several vital physiologic parameters (e.g., ECG, blood pressure, heart rate, temperature, cardiac output, apnoea, and respiratory/anaesthetic gas concentrations) of one patient in intensive or general healthcare settings. It typically set-up by the patient bedside and often includes programmable alarms, portable radio transmitters, receivers, and antennas (telemetry systems) to allow monitoring during patient ambulation or transportation; it is not dedicated to neonatal use and is not intended to be worn on the body.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Low exhaled tidal volume, high leak alarm, inability to maintain PEEP.
Root cause: Diaphragm/valve degraded by secretions and cleaning cycles.
Symptoms: Inaccurate delivered volume, false volume/pressure alarms.
Root cause: Sensor contamination or aging of differential-pressure transducer.
Symptoms: Unexpected shutdown, persistent low-battery alarm.
Root cause: Aged internal battery or failed PSU.
Symptoms: FiO2 out of range, oxygen alarm.
Root cause: Failed solenoid or proportional valve in blender.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable - replace per PM schedule.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Flow / differential pressure sensor | Periodic replacement | $$$ | Moderate |
| Bacterial / HEPA filter Consumable - replace per PM schedule. | Consumable | $ | Easy |
| Compressor / turbine module | Periodic replacement | $$$ | Difficult |
| Rechargeable battery pack | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
A defective electrical component in the battery charging circuit in two lots of the ProBP 2400 may potentially cause an over-voltage battery charging condition when the device is connected to the external power supply/battery charger. This potential defect resides with the device, not the battery. The over-voltage condition can result in high NiMH battery temperatures that, in some cases, can reach levels sufficient to cause melting of the plastic battery door and other adjacent plastic and foam device components.
The firm became aware that the KleenSpec 788 Corded Illuminator has the potential to be incorrectly connected to a commonly available AC power cord instead of the illuminator low voltage transformer power cord which is supplied with the illuminator. This may result in a shattered lamp bulb and a potential electric shock hazard.
GMP deficiencies
Incorrect location of inflatable portion of cuff that may result in cuff falling off patient's arm, error code when used with patient monitor, or an inaccurate reading.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →