Verathon · Model BladderScan Prime Plus
ImageSense deep learning technology automatically calculates 3D volume Enhanced EdgeScan off-center scan alerts Pioneering pubic bone interference indicator 12 BMODE-slice review Time saving, one-step automated volume results BladderTraq precision and 1–999 mL volume detection Automatically adapts to patient types such as male, female, and small child Unique probe design Cart adaptable to a variety of clinical set-ups Password-protected patient information No sonographer required Fully sealed probe design which protects sensitive electronics against accidental moisture spills Maximized up-time with two replaceable batteries; no console charging required Expanded list of cleaners and disinfectants for a variety of hospital requirements.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No scan on the button press, implausible bladder volume, aiming reticle unstable.
Root cause: Probe / transducer fault, a probe-cable fault, or a calibration issue.
Symptoms: Short run time, will not turn on, will not charge in the cradle.
Root cause: Aged battery or a charging-contact / cradle fault.
Symptoms: Image drops out when the cable is moved, probe not detected.
Root cause: Broken conductors in the probe cable or a worn connector.
Symptoms: Blank or frozen display, touchscreen unresponsive.
Root cause: Display or console-board fault.
Symptoms: No or faint printout.
Root cause: Thermal printer or paper.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Probe / transducer | Periodic replacement | $$$ | Moderate |
| Battery | Periodic replacement | $$ | Easy |
| Probe cable / connector | Periodic replacement | $$ | Fairly easy |
| Display / console board | Major service part | $$$ | Moderate |
| Thermal printer / paper | Periodic replacement | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Due to a change in handle material and process, probe handle may crack.