Verathon · Model BladderScan BVI 6400
Helps assess urinary retention and postoperative urinary retention (POUR) Helps prevent unnecessary catheterization Helps reduce rates of catheter-associated urinary tract infection (CAUTI) Helps differentiate between types of incontinence to determine appropriate care Helps monitor postoperative recovery Helps improve efficiency of healthcare professionals by reducing costs and saving staff time.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No scan on the button press, implausible bladder volume, aiming reticle unstable.
Root cause: Probe / transducer fault, a probe-cable fault, or a calibration issue.
Symptoms: Short run time, will not turn on, will not charge in the cradle.
Root cause: Aged battery or a charging-contact / cradle fault.
Symptoms: Image drops out when the cable is moved, probe not detected.
Root cause: Broken conductors in the probe cable or a worn connector.
Symptoms: Blank or frozen display, touchscreen unresponsive.
Root cause: Display or console-board fault.
Symptoms: No or faint printout.
Root cause: Thermal printer or paper.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Probe / transducer | Periodic replacement | $$$ | Moderate |
| Battery | Periodic replacement | $$ | Easy |
| Probe cable / connector | Periodic replacement | $$ | Fairly easy |
| Display / console board | Major service part | $$$ | Moderate |
| Thermal printer / paper | Periodic replacement | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
The firm is providing customers with an updated Operations and Maintenance Manual for each of the BladderScan BVI 9600 and AortaScan AMI 9700 devices. The updated Operations and Maintenance Manuals clarify that these devices should not be used for the screening, detection, or diagnosis of abdominal aortic aneurysms (AAAs).
The BladderScan BVI 9600 with AortaScan devices may experience the loss of a factory installed software configuration file which results in a loss of the aorta measurement function.
Three specific issues are being addressed: 1. Calculation and use of an incorrect year in dates after 12/31/2009. 2. Double scans, continuous scans, and double printing. 3. Calibration sensitivity.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.